Instrument knowledge|Precautions for installation of electromagnetic flowmeter

DATE: 2020-08-06
BROWSE TIMES: 13

Instrument knowledge|Precautions for installation of electromagnetic flowmeter

 

Electromagnetic flow meters can measure conductive liquids and are widely used in chemical, chemical fiber, food, paper, sugar, mining and metallurgy, urban water supply and drainage, environmental protection, water conservancy, steel, petroleum, pharmaceuticals and other industrial fields to measure various acids , Alkali, salt solution, mud, mineral pulp, paper pulp, coal water slurry, corn steep liquor, fiber pulp, grain pulp, lime milk, sewage, etc.。

Instrument knowledge|Precautions for installation of electromagnetic flowmeter

Electromagnetic flowmeter has a series of excellent characteristics, which can solve the problems that other flowmeters are not easy to use, such as the measurement of dirty flow and corrosive flow.

 

Advantage

The measurement channel is a smooth straight pipe, which will not be blocked, and is suitable for measuring liquid-solid two-phase fluids containing solid particles, such as pulp, mud, sewage, etc.;

The detection will not cause pressure loss, and the energy saving effect is good;

The measured volume flow is actually not significantly affected by changes in fluid density, viscosity, temperature, pressure and conductivity;

Wide range of turndown ratio and wide caliber range;

Corrosive fluids can be used.

 

Disadvantage

Cannot be used to measure gas, steam and liquids containing a large amount of gas;

It cannot be used to measure liquid media with very low conductivity, and the electromagnetic flowmeter cannot measure petroleum products or organic solvents and other media;

It cannot be used to measure high temperature media. At present, it is restricted by the lining material and electrical insulation material of the measuring tube of the electromagnetic flowmeter for general industry;

Electromagnetic flowmeters are easily affected by external electromagnetic interference.

 

Installation Precautions

In order to make the electromagnetic flowmeter meet the working conditions of the measurement site and meet the accuracy and function requirements of the instrument, the following precautions are required during the installation process:


1. Install according to the calibration direction

The direction of the arrow on the sensor must be consistent with the actual flow direction. (If the installation is reversed, the flow direction item in the converter menu must be corrected)


2. Ensure that the full pipe is installed

The full tube measurement state must be ensured, and the non-full tube state cannot occur.

Instrument knowledge|Precautions for installation of electromagnetic flowmeter

Avoid installation upstream of the downward pipeline;

Avoid installing at the highest point of the pipeline, which is easy to collect bubbles;

Instrument knowledge|Precautions for installation of electromagnetic flowmeter

When installing on a horizontal pipeline, the two measuring electrodes should not be directly above and directly below the pipeline. When installed vertically, the fluid should flow from bottom to top.

 

3.Straight pipe distance requirements

Must ensure that the first 10 and rear 5 straight pipe sections.

Instrument knowledge|Precautions for installation of electromagnetic flowmeter


4.Series and parallel installation

If several sensors need to be connected in series on the same pipe in sequence, the distance between each sensor should be at least the length of 2 sensors. If two or more sensors are installed parallel to each other, the distance between the sensors must be greater than 1m.

5. installation and bypass pipe

The pipeline where the flowmeter is installed should have a valve that can cut off the medium in the pipeline to facilitate daily maintenance and maintenance. When the process does not allow flow interruption. A bypass pipe can be added to the pipe where the flowmeter is installed.

6.Sensor ground

The sensor must be well grounded separately


Grounding meaning:

Give the signal source a reference to zero potential, so that the measured medium is connected to the earth. The safety requirements (such as lightning strikes) are against external interference. The grounding point is as close as possible to the sensor, and it must not be connected to the grounding point of other equipment. Ground resistance ≤10Ω

 

Grounding requirements:

Electromagnetic flowmeter transmitter housing, shielded wire, measuring conduit and pipes at both ends of the transmitter must be grounded and set a separate grounding point, and must not be connected to public ground wires or upper and lower pipes such as electric motors; the converter part has passed the cable Line grounding, there is no need to ground again to avoid interference due to different potentials.

 

Grounding method with grounding ring:

If the process pipeline is made of insulating material, it must be equipped with a grounding ring to stabilize the boundary conditions. Improve the measurement qualitative and accuracy.

 

7.Ensure that the split type electromagnetic flowmeter is matched and installed

The split type electromagnetic flowmeter should ensure that the electromagnetic flow converter and the sensor are matched with the number, and the number of the converter and the matching sensor is the same (Zhang Guan Li Dai will not work normally)


8. Wire and lay cables correctly

Must be wired and cabled correctly. The connecting cables of the sensor and the converter must be seated in correct numbers and connected correctly (understand the meaning of each line number)

 

Cable installation precautions: The signal cable and the excitation cable are separated through pipes (iron pipes) to prevent interference and mechanical damage. The special cable provided by the company must be used. The signal cable, excitation cable and power line must be installed separately, and the pulse output line should be considered between the lines. Distributed capacitance should not be too long.

 

9 .Reasonable configuration of the rear position instrument

The rear instrument of the electromagnetic flowmeter must be reasonably configured

The active or passive output should be determined according to the requirements of the subsequent instrument. The external power supply for passive output should meet the technical specifications. If the resistance is connected in series, the resistance value should be within the allowable range.

 

10 .proofreading the power supply voltage

The power supply voltage must be calibrated before commissioning (the phenomenon of 24VDC instrument being burned out by 220VAC power supply is endless)

During debugging, only zero position adjustment is allowed, and various parameters are set after factory calibration without any need to change.

 


Share to
RELATED NEWS / News More
  • Number of clicks: 30
    2021 - 04 - 08
    The 80Ghz radar level meter is used in the measurement of organic liquids. The reactor contains multiple layers of stirring and has a small range; after repeated tests, the 80Ghz FMCW radar level meter has always had a stable measurement effect and accurate data, which has won the satisfaction of end users .
  • Number of clicks: 1
    2021 - 04 - 06
    1. 1 Description of productsElectromagnetic flow meter is a kind of inductional instrument designed by Faradays' law of electromagnetic induction to measure flow of conductive media in the tube. It adopts the technology of insertion of single chip to realize digital excitation and employs CAN local fieldbus.Electromagnetic flow meter can realize local indication and output electrical current signal of 4-20mA which can be used to record, adjust and control. Electromagnetic flow meters are widely used in industrial sectors such as chemical industry, environmental protection, metallurgy, pharmaceutical, paper making, water supply and removal etc.Besides measuring flow of general conductive liquid electromagnetic flow meter can measure flow of liquid-solid fluid, high-viscosity fluid ...
  • Number of clicks: 101
    2021 - 04 - 01
    Common problems of radar level transmitter Radar level transmitter are mainly installed in various metal and non-metal containers or pipelines to perform non-contact continuous measurement of the level of liquids, slurries and particles. It is also suitable for occasions with dust, inert gas and steam. The radar level transmitter has the characteristics of safety, cleanliness, high precision, long life, stable and reliable, convenient installation and maintenance, and simple reading. Because there are no moving parts, maintenance is minimal, but some minor faults will inevitably occur during use. Here is a breakdown analysis and countermeasures for everyone.1. If an alarm occurs when the liquid level is within the applicable range, then:a. When the liquid leve...
  • Number of clicks: 193
    2021 - 03 - 30
    Radar level meter is a common product type of level meter, which has the advantages of accurate measurement, stable performance, high reliability, easy maintenance, and wide application range. It can be installed in various metal and non-metal containers or tanks to measure the level of liquid, slurry and granular materials. So, under what special conditions will it be installed?1. Selection under the condition of stirringIn a tank with stirring, guided wave radar level gauges are generally not used. The strength of stirring can easily bring the cable of the guided wave radar into the stirring device, or bend the rod of the guided wave radar. So you can only choose non-contact measurement of FM or high-frequency horn-mouth radar level meter to measure.2. Selection under vacuum conditionsUn...
  • Number of clicks: 103
    2021 - 03 - 25
    Radar level meter measures ammonia in a fertilizer plant Ammonia refers to an aqueous solution of ammonia with a strong pungent odor and weak alkaline. Store in a cool, dry, ventilated place, away from fire and heat. Avoid direct sunlight and keep the container tightly closed. It should be stored separately from acids, metal powders, etc. Open-air storage tanks must have cooling measures in summer. Ammonia is often used in agricultural fertilizers, in the chemical industry for the manufacture of various ammonium salts, aminating agents for organic synthesis, and catalysts for the production of thermosetting phenolic resins. In the textile industry, it is used in wool spinning, silk, printing and dyeing industries, for washing wool, woolen cloth, grey cloth, oil stains, dyein...
Email:sales@dechuang-auto.com
Adress:
Tel:86-29 88606479
Fax: 86-29 88385903
B203, Gazelle Valley, No.69 Jinye Rd, Hi-tech zone, Xi'an, Shaanxi Province, P.R.China. 710077
Copyright ©2018 - 2019 Dechuang Electrical Technology Co., Ltd SITEMAP
Rhino Cloud Provides Enterprise Cloud Services
犀牛云提供云计算服务